Line 6: | Line 6: |
* [[KAGRA/Commissioning/Phase1/Operation/MeasurementsSummary|Phase-1 Measurements Summary]] | * [[KAGRA/Commissioning/Phase1/Operation/MeasurementsSummary|Phase 1 Measurements Summary]] See following pages for some design parameters * [[KAGRA/Subgroups/IOO/OptParam|IOO related]] * [[KAGRA/Subgroups/IOO/OptParam|MIF related]] == RF Sidebands == * f1 frequency: 16.8732705 MHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831 * f2 frequency: 44.995388 MHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831 * f1 modulation depth: * f2 modulation depth: |
Line 9: | Line 18: |
* Length | * Round-trip length * '''Design''': 53.28708 m (FSR of 5.6269873 MHz) * '''Measured''': 53.302985(9) m B) * '''Measurement method''': FSR was measured to be 5.624308(1) MHz. https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831 * '''Notes''': Measured on April 2018. Any re-installation after this? Needs to be adjusted to match with PRC length. * Finesse * '''Design''': 520 * '''Measured''': 409 * '''Measurement method''': Cavity pole was measured to be 6.86(2) kHz. https://klog.icrr.u-tokyo.ac.jp/osl/?r=14569 * '''Notes''': Lower than iKAGRA measurement of 540+/-70 http://klog.icrr.u-tokyo.ac.jp/osl/?r=618 Degraded? :o * Mode-match * '''Measured''': * '''Measurement method''': * '''Notes''': == Arm Cavities == * X-arm length * '''Design''': 3000 m * '''Measured''': 2999.990(2) m B) * '''Measurement method''': Cavity scan FSR = 49.96557(4) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332 * '''Notes''': * X-arm Gouy phase * '''Design''': 250.75 deg * '''Measured''': 250.7(3) deg * '''Measurement method''': Cavity scan TMS=34.79(5) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332 * '''Notes''': * X-arm finesse for IR * '''Design''': 1530 * '''Measured''': 1410(30) * '''Measurement method''': Ring down * '''Notes''': There is temperature dependence due to frosting :o https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033 * X-arm round-trip loss * '''Design''': 100 ppm * '''Measured''': 86(3) ppm * '''Measurement method''': From reflectivity change in locked and unlocked https://klog.icrr.u-tokyo.ac.jp/osl/?r=7307 * '''Notes''': The estimate might be wrong due to birefringence https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=11633 :o * X-arm mode-match * '''Measured''': 91(1)% * '''Measurement method''': Cavity scan https://klog.icrr.u-tokyo.ac.jp/osl/?r=7307 * '''Notes''': * X-arm finesse for green * '''Design''': 51 * '''Measured''': 41(2) * '''Measurement method''': Calculated from reflectivity measurements by Hirose-san * '''Notes''': * Y-arm length * '''Design''': 3000 m * '''Measured''': 3000.015(1) m B) * '''Measurement method''': Cavity scan, FSR = 49.96517(2) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505 * '''Notes''': * Y-arm Gouy phase * '''Design''': 250.75 deg * '''Measured''': 250.9(3) deg * '''Measurement method''': Cavity scan, TMS = 34.82(5) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505 * '''Notes''': * Y-arm finesse for IR * '''Design''': 1530 * '''Measured''': 1320(30) * '''Measurement method''': Ring down, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9052, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9211, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9496 * '''Notes''': There is temperature dependence due to frosting :o https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033 * Y-arm round-trip loss * '''Design''': 100 ppm * '''Measured''': NOT YET MEASURED? :o * '''Measurement method''': * '''Notes''': * Y-arm mode-match * '''Measured''': 87(2) % * '''Measurement method''': Cavity scan https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505 * '''Notes''': * Y-arm finesse for green * '''Design''': 51 * '''Measured''': 53(3) * '''Measurement method''': Calculated from reflectivity measurements by Hirose-san * '''Notes''': == Power Recycling Cavity == * PRX length * '''Design''': 68.2563 m (PRC length = 66.591 m) * '''Measured''': 68 m * '''Measurement method''': Using dnu/nu=dL/L https://klog.icrr.u-tokyo.ac.jp/osl/?r=14633 * '''Notes''': Compare it with PRY. * PRY length * '''Design''': 64.9265 m * '''Measured''': 70.2 m * '''Measurement method''': Using dnu/nu=dL/L https://klog.icrr.u-tokyo.ac.jp/osl/?r=14565 * '''Notes''': In consistent results between different methods and not compatible with Schnupp asymmetry measurement :o * PRC Gouy phase * '''Design''': 33.0 deg * '''Measured''': 31(1) deg * '''Measurement method''': Cavity scan with the inertia locking method http://klog.icrr.u-tokyo.ac.jp/osl/?r=9276 * '''Notes''': After PR2 was moved. == Signal Recycling Cavity == * SRX length * '''Design''': 68.2563 m (SRC length = 66.591 m) * '''Measured''': * '''Measurement method''': * '''Notes''': * SRY length * '''Design''': 64.9265 m * '''Measured''': * '''Measurement method''': * '''Notes''': * SRC Gouy phase * '''Design''': 35.0 * '''Measured''': 48(2) :o * '''Measurement method''': Cavity scan with the inertia locking method http://klog.icrr.u-tokyo.ac.jp/osl/?r=9246 * '''Notes''': == Michelson == * Schnupp asymmetry * '''Design''': 3.3299 m * '''Measured''': 3.36(1) m * '''Measurement method''': Demodulation phase difference in locking X-arm and Y-arm https://klog.icrr.u-tokyo.ac.jp/osl/?r=14575 * '''Notes''': Not consistent with PRX and PRY measurements. == Dual Recycled Michelson == * Power recycling gain * '''Design''': * '''Measured''': ~3 * '''Measurement method''': REFERENCE TO BE ADDED * '''Notes''': * Power at each ISC detection ports when DRFPMI is locked * IMC output: * REFL: * POP p-pol: * POP s-pol: * POS p-pol: * POS s-pol: * AS RF: * OMC REFL: * OMC TRANS: * TRX: * TRY: == Power Recycled Fabry-Perot Michelson == * Power recycling gain * '''Design''': 10 * '''Measured''': 11-12 * '''Measurement method''': REFERENCE TO BE ADDED * '''Notes''': * Power at each ISC detection ports when PRFPMI is locked * IMC output: * REFL: * POP p-pol: * POP s-pol: * POS p-pol: * POS s-pol: * AS RF: * OMC REFL: * OMC TRANS: * TRX: * TRY: == Output Mode Cleaner == * Round-trip length |
Line 15: | Line 179: |
== Arm Cavities == == Power Recycling Cavity == == Signal Recycling Cavity == == Michelson == == Dual Recycled Michelson == == Power Recycled Fabry-Perot Michelson == == Output Mode Cleaner == |
* '''Design''': * '''Measured''': * '''Measurement method''': * '''Notes''': |
Optical Characterization
This page is intended for summarizing the interferometer optical characterization results and discussions. See following pages and documents for iKAGRA and bKAGRA Phase-1 characterization summary:
See following pages for some design parameters
RF Sidebands
f1 frequency: 16.8732705 MHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831
f2 frequency: 44.995388 MHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831
- f1 modulation depth:
- f2 modulation depth:
Input Mode Cleaner
- Round-trip length
Design: 53.28708 m (FSR of 5.6269873 MHz)
Measured: 53.302985(9) m
Measurement method: FSR was measured to be 5.624308(1) MHz. https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831
Notes: Measured on April 2018. Any re-installation after this? Needs to be adjusted to match with PRC length.
- Finesse
Design: 520
Measured: 409
Measurement method: Cavity pole was measured to be 6.86(2) kHz. https://klog.icrr.u-tokyo.ac.jp/osl/?r=14569
Notes: Lower than iKAGRA measurement of 540+/-70 http://klog.icrr.u-tokyo.ac.jp/osl/?r=618 Degraded?
- Mode-match
Measured:
Measurement method:
Notes:
Arm Cavities
- X-arm length
Design: 3000 m
Measured: 2999.990(2) m
Measurement method: Cavity scan FSR = 49.96557(4) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332
Notes:
- X-arm Gouy phase
Design: 250.75 deg
Measured: 250.7(3) deg
Measurement method: Cavity scan TMS=34.79(5) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332
Notes:
- X-arm finesse for IR
Design: 1530
Measured: 1410(30)
Measurement method: Ring down
Notes: There is temperature dependence due to frosting
https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033
- X-arm round-trip loss
Design: 100 ppm
Measured: 86(3) ppm
Measurement method: From reflectivity change in locked and unlocked https://klog.icrr.u-tokyo.ac.jp/osl/?r=7307
Notes: The estimate might be wrong due to birefringence https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=11633
- X-arm mode-match
Measured: 91(1)%
Measurement method: Cavity scan https://klog.icrr.u-tokyo.ac.jp/osl/?r=7307
Notes:
- X-arm finesse for green
Design: 51
Measured: 41(2)
Measurement method: Calculated from reflectivity measurements by Hirose-san
Notes:
- Y-arm length
Design: 3000 m
Measured: 3000.015(1) m
Measurement method: Cavity scan, FSR = 49.96517(2) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505
Notes:
- Y-arm Gouy phase
Design: 250.75 deg
Measured: 250.9(3) deg
Measurement method: Cavity scan, TMS = 34.82(5) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505
Notes:
- Y-arm finesse for IR
Design: 1530
Measured: 1320(30)
Measurement method: Ring down, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9052, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9211, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9496
Notes: There is temperature dependence due to frosting
https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033
- Y-arm round-trip loss
Design: 100 ppm
Measured: NOT YET MEASURED?
Measurement method:
Notes:
- Y-arm mode-match
Measured: 87(2) %
Measurement method: Cavity scan https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505
Notes:
- Y-arm finesse for green
Design: 51
Measured: 53(3)
Measurement method: Calculated from reflectivity measurements by Hirose-san
Notes:
Power Recycling Cavity
- PRX length
Design: 68.2563 m (PRC length = 66.591 m)
Measured: 68 m
Measurement method: Using dnu/nu=dL/L https://klog.icrr.u-tokyo.ac.jp/osl/?r=14633
Notes: Compare it with PRY.
- PRY length
Design: 64.9265 m
Measured: 70.2 m
Measurement method: Using dnu/nu=dL/L https://klog.icrr.u-tokyo.ac.jp/osl/?r=14565
Notes: In consistent results between different methods and not compatible with Schnupp asymmetry measurement
- PRC Gouy phase
Design: 33.0 deg
Measured: 31(1) deg
Measurement method: Cavity scan with the inertia locking method http://klog.icrr.u-tokyo.ac.jp/osl/?r=9276
Notes: After PR2 was moved.
Signal Recycling Cavity
- SRX length
Design: 68.2563 m (SRC length = 66.591 m)
Measured:
Measurement method:
Notes:
- SRY length
Design: 64.9265 m
Measured:
Measurement method:
Notes:
- SRC Gouy phase
Design: 35.0
Measured: 48(2)
Measurement method: Cavity scan with the inertia locking method http://klog.icrr.u-tokyo.ac.jp/osl/?r=9246
Notes:
Michelson
- Schnupp asymmetry
Design: 3.3299 m
Measured: 3.36(1) m
Measurement method: Demodulation phase difference in locking X-arm and Y-arm https://klog.icrr.u-tokyo.ac.jp/osl/?r=14575
Notes: Not consistent with PRX and PRY measurements.
Dual Recycled Michelson
- Power recycling gain
Design:
Measured: ~3
Measurement method: REFERENCE TO BE ADDED
Notes:
- Power at each ISC detection ports when DRFPMI is locked
- IMC output:
- REFL:
- POP p-pol:
- POP s-pol:
- POS p-pol:
- POS s-pol:
- AS RF:
- OMC REFL:
- OMC TRANS:
- TRX:
- TRY:
Power Recycled Fabry-Perot Michelson
- Power recycling gain
Design: 10
Measured: 11-12
Measurement method: REFERENCE TO BE ADDED
Notes:
- Power at each ISC detection ports when PRFPMI is locked
- IMC output:
- REFL:
- POP p-pol:
- POP s-pol:
- POS p-pol:
- POS s-pol:
- AS RF:
- OMC REFL:
- OMC TRANS:
- TRX:
- TRY:
Output Mode Cleaner
- Round-trip length
Design:
Measured:
Measurement method:
Notes:
- Finesse
Design:
Measured:
Measurement method:
Notes: