Size: 3288
Comment:
|
Size: 5528
Comment:
|
Deletions are marked like this. | Additions are marked like this. |
Line 19: | Line 19: |
* '''Design''': * '''Measured''': * '''Measurement method''': * '''Notes''': |
* '''Design''': 53.28708 m (FSR of 5.6269873 MHz) * '''Measured''': 53.302985(9) m B) * '''Measurement method''': FSR was measured to be 5.624308(1) MHz. https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831 * '''Notes''': Measured on April 2018. Any re-installation after this? |
Line 24: | Line 24: |
* '''Design''': | * '''Design''': ~500 * '''Measured''': 409 * '''Measurement method''': Cavity pole was measured to be 6.86(2) kHz. https://klog.icrr.u-tokyo.ac.jp/osl/?r=14569 * '''Notes''': Lower than iKAGRA measurement of 540+/-70 http://klog.icrr.u-tokyo.ac.jp/osl/?r=618 Degraded? * Mode-match |
Line 31: | Line 35: |
* '''Design''': * '''Measured''': * '''Measurement method''': |
* '''Design''': 3000 m * '''Measured''': 2999.990(2) m * '''Measurement method''': Cavity scan FSR = 49.96557(4) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332 * '''Notes''': * X-arm Gouy phase * '''Design''': 250.75 deg * '''Measured''': 250.7(3) deg * '''Measurement method''': Cavity scan TMS=34.79(5) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332 |
Line 36: | Line 45: |
* '''Design''': * '''Measured''': * '''Measurement method''': * '''Notes''': |
* '''Design''': 1530 * '''Measured''': 1410(30) * '''Measurement method''': Ring down * '''Notes''': There is temperature dependence due to frosting :o https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033 |
Line 41: | Line 50: |
* '''Design''': | * '''Design''': 100 ppm * '''Measured''': 86(3) ppm * '''Measurement method''': From reflectivity change in locked and unlocked https://klog.icrr.u-tokyo.ac.jp/osl/?r=7307 * '''Notes''': The estimate might be wrong due to birefringence https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=11633 :o * X-arm mode-match |
Line 46: | Line 59: |
* '''Design''': | * '''Design''': 51 * '''Measured''': 41(2) * '''Measurement method''': Calculated from reflectivity measurements by Hirose-san * '''Notes''': * Y-arm length * '''Design''': 3000 m * '''Measured''': 3000.015(1) m B) * '''Measurement method''': Cavity scan, FSR = 49.96517(2) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505 * '''Notes''': * Y-arm Gouy phase * '''Design''': 250.75 deg * '''Measured''': 250.9(3) deg * '''Measurement method''': Cavity scan, TMS = 34.82(5) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505 * '''Notes''': * Y-arm finesse for IR * '''Design''': 1530 * '''Measured''': 1320(30) * '''Measurement method''': Ring down, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9052, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9211, http://klog.icrr.u- tokyo.ac.jp/osl/?r=9496 * '''Notes''': There is temperature dependence due to frosting :o https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033 * Y-arm round-trip loss * '''Design''': 100 ppm |
Line 50: | Line 83: |
* Y-arm length * '''Design''': * '''Measured''': * '''Measurement method''': * '''Notes''': * Y-arm finesse for IR * '''Design''': * '''Measured''': * '''Measurement method''': * '''Notes''': * Y-arm round-trip loss * '''Design''': |
* Y-arm mode-match |
Line 66: | Line 88: |
* '''Design''': * '''Measured''': * '''Measurement method''': |
* '''Design''': 51 * '''Measured''': 53(3) * '''Measurement method''': Calculated from reflectivity measurements by Hirose-san |
Line 82: | Line 104: |
* PRC Gouy phase * '''Design''': * '''Measured''': * '''Measurement method''': * '''Notes''': |
|
Line 90: | Line 117: |
* '''Design''': * '''Measured''': * '''Measurement method''': * '''Notes''': * SRC Gouy phase |
Optical Characterization
This page is intended for summarizing the interferometer optical characterization results and discussions. See following pages and documents for iKAGRA and bKAGRA Phase-1 characterization summary:
See following pages for some design parameters
RF Sidebands
- f1 frequency
- f2 frequency
- f1 modulation depth
- f2 modulation depth
Input Mode Cleaner
- Round-trip length
Design: 53.28708 m (FSR of 5.6269873 MHz)
Measured: 53.302985(9) m
Measurement method: FSR was measured to be 5.624308(1) MHz. https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831
Notes: Measured on April 2018. Any re-installation after this?
- Finesse
Design: ~500
Measured: 409
Measurement method: Cavity pole was measured to be 6.86(2) kHz. https://klog.icrr.u-tokyo.ac.jp/osl/?r=14569
Notes: Lower than iKAGRA measurement of 540+/-70 http://klog.icrr.u-tokyo.ac.jp/osl/?r=618 Degraded?
- Mode-match
Measured:
Measurement method:
Notes:
Arm Cavities
- X-arm length
Design: 3000 m
Measured: 2999.990(2) m
Measurement method: Cavity scan FSR = 49.96557(4) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332
Notes:
- X-arm Gouy phase
Design: 250.75 deg
Measured: 250.7(3) deg
Measurement method: Cavity scan TMS=34.79(5) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332
Notes:
- X-arm finesse for IR
Design: 1530
Measured: 1410(30)
Measurement method: Ring down
Notes: There is temperature dependence due to frosting
https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033
- X-arm round-trip loss
Design: 100 ppm
Measured: 86(3) ppm
Measurement method: From reflectivity change in locked and unlocked https://klog.icrr.u-tokyo.ac.jp/osl/?r=7307
Notes: The estimate might be wrong due to birefringence https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=11633
- X-arm mode-match
Measured:
Measurement method:
Notes:
- X-arm finesse for green
Design: 51
Measured: 41(2)
Measurement method: Calculated from reflectivity measurements by Hirose-san
Notes:
- Y-arm length
Design: 3000 m
Measured: 3000.015(1) m
Measurement method: Cavity scan, FSR = 49.96517(2) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505
Notes:
- Y-arm Gouy phase
Design: 250.75 deg
Measured: 250.9(3) deg
Measurement method: Cavity scan, TMS = 34.82(5) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505
Notes:
- Y-arm finesse for IR
Design: 1530
Measured: 1320(30)
Measurement method: Ring down, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9052, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9211, http://klog.icrr.u- tokyo.ac.jp/osl/?r=9496
Notes: There is temperature dependence due to frosting
https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033
- Y-arm round-trip loss
Design: 100 ppm
Measured:
Measurement method:
Notes:
- Y-arm mode-match
Measured:
Measurement method:
Notes:
- Y-arm finesse for green
Design: 51
Measured: 53(3)
Measurement method: Calculated from reflectivity measurements by Hirose-san
Notes:
Power Recycling Cavity
- PRX length
Design:
Measured:
Measurement method:
Notes:
- PRY length
Design:
Measured:
Measurement method:
Notes:
- PRC Gouy phase
Design:
Measured:
Measurement method:
Notes:
Signal Recycling Cavity
- SRX length
Design:
Measured:
Measurement method:
Notes:
- SRY length
Design:
Measured:
Measurement method:
Notes:
- SRC Gouy phase
Design:
Measured:
Measurement method:
Notes:
Michelson
- Schnupp asymmetry
Design:
Measured:
Measurement method:
Notes:
Dual Recycled Michelson
- Power recycling gain
Design:
Measured:
Measurement method:
Notes:
Power Recycled Fabry-Perot Michelson
- Power recycling gain
Design:
Measured:
Measurement method:
Notes:
- Power at each ISC detection ports when locked
- REFL:
- POP p-pol:
- POP s-pol:
- POS p-pol:
- POS s-pol:
- AS RF:
- OMC REFL:
- OMC TRANS:
- TRX:
- TRY:
Output Mode Cleaner
- Round-trip length
Design:
Measured:
Measurement method:
Notes:
- Finesse
Design:
Measured:
Measurement method:
Notes: