Line 12: Line 12:
 * f1 frequency
 * f2 frequency
 * f1 modulation depth
 * f2 modulation depth
 * f1 frequency: 16.8732705 MHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831
 * f2 frequency: 44.995388 MHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=4831
 * f1 modulation depth:
 * f2 modulation depth:
Line 22: Line 22:
  * '''Notes''': Measured on April 2018. Any re-installation after this?   * '''Notes''': Measured on April 2018. Any re-installation after this? Needs to be adjusted to match with PRC length.
Line 24: Line 24:
  * '''Design''': ~500   * '''Design''': 520
Line 27: Line 27:
  * '''Notes''': Lower than iKAGRA measurement of 540+/-70 http://klog.icrr.u-tokyo.ac.jp/osl/?r=618   * '''Notes''': Lower than iKAGRA measurement of 540+/-70 http://klog.icrr.u-tokyo.ac.jp/osl/?r=618 Degraded? :o
Line 35: Line 35:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Design''': 3000 m
  * '''Measured''': 2999.990(2) m B)
  * '''Measurement method''': Cavity scan FSR = 49.96557(4) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332
  * '''Notes''':
 * X-arm Gouy phase
  * '''Design''': 250.75 deg
  * '''Measured''': 250.7(3) deg
  * '''Measurement method''': Cavity scan TMS=34.79(5) kHz http://klog.icrr.u-tokyo.ac.jp/osl/?r=7332
Line 40: Line 45:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Notes''':
  * '''Design''': 1530
  * '''Measured''': 1410(30)
  * '''Measurement method''': Ring down
  * '''Notes''': There is temperature dependence due to frosting :o https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033
Line 45: Line 50:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Notes''':
  * '''Design''': 100 ppm
  * '''Measured''': 86(3) ppm
  * '''Measurement method''': From reflectivity change in locked and unlocked https://klog.icrr.u-tokyo.ac.jp/osl/?r=7307
  * '''Notes''': The estimate might be wrong due to birefringence https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=11633 :o
Line 50: Line 55:
  * '''Measured''':
  * '''Measurement method''':
  * '''Measured''': 91(1)%
  * '''Measurement method''': Cavity scan https://klog.icrr.u-tokyo.ac.jp/osl/?r=7307
Line 54: Line 59:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Design''': 51
  * '''Measured''': 41(2)
  * '''Measurement method''': Calculated from reflectivity measurements by Hirose-san
Line 59: Line 64:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Design''': 3000 m
  * '''Measured''': 3000.015(1) m B)
  * '''Measurement method''': Cavity scan, FSR = 49.96517(2) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505
  * '''Notes''':
 * Y-arm Gouy phase
  * '''Design''': 250.75 deg
  * '''Measured''': 250.9(3) deg
  * '''Measurement method''': Cavity scan, TMS = 34.82(5) kHz https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505
Line 64: Line 74:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Notes''':
  * '''Design''': 1530
  * '''Measured''': 1320(30)
  * '''Measurement method''': Ring down, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9052, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9211, http://klog.icrr.u-tokyo.ac.jp/osl/?r=9496
  * '''Notes''': There is temperature dependence due to frosting :o https://klog.icrr.u-tokyo.ac.jp/osl/?r=10033
Line 69: Line 79:
  * '''Design''':
  * '''Measured''':
  * '''Design''': 100 ppm
  * '''Measured''': NOT YET MEASURED? :o
Line 74: Line 84:
  * '''Measured''':
  * '''Measurement method''':
  * '''Measured''': 87(2) %
  * '''Measurement method''': Cavity scan https://klog.icrr.u-tokyo.ac.jp/osl/?r=9505
Line 78: Line 88:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Design''': 51
  * '''Measured''': 53(3)
  * '''Measurement method''': Calculated from reflectivity measurements by Hirose-san
Line 85: Line 95:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Notes''':
  * '''Design''': 68.2563 m (PRC length = 66.591 m)
  * '''Measured''': 68 m
  * '''Measurement method''': Using dnu/nu=dL/L https://klog.icrr.u-tokyo.ac.jp/osl/?r=14633
  * '''Notes''': Compare it with PRY.
Line 90: Line 100:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Notes''':
  * '''Design''': 64.9265 m
  * '''Measured''': 70.2 m
  * '''Measurement method''': Using dnu/nu=dL/L https://klog.icrr.u-tokyo.ac.jp/osl/?r=14565
  * '''Notes''': In consistent results between different methods and not compatible with Schnupp asymmetry measurement :o
Line 95: Line 105:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Notes''':
  * '''Design''': 33.0 deg
  * '''Measured''': 31(1) deg
  * '''Measurement method''': Cavity scan with the inertia locking method http://klog.icrr.u-tokyo.ac.jp/osl/?r=9276
  * '''Notes''': After PR2 was moved.
Line 102: Line 112:
  * '''Design''':   * '''Design''': 68.2563 m (SRC length = 66.591 m)
Line 107: Line 117:
  * '''Design''':   * '''Design''': 64.9265 m
Line 112: Line 122:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Design''': 35.0
  * '''Measured''': 48(2) :o
  * '''Measurement method''': Cavity scan with the inertia locking method http://klog.icrr.u-tokyo.ac.jp/osl/?r=9246
Line 119: Line 129:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Notes''':
  * '''Design''': 3.3299 m
  * '''Measured''': 3.36(1) m
  * '''Measurement method''': Demodulation phase difference in locking X-arm and Y-arm https://klog.icrr.u-tokyo.ac.jp/osl/?r=14575
  * '''Notes''': Not consistent with PRX and PRY measurements.
Line 127: Line 137:
  * '''Measured''':
  * '''Measurement method''':
  * '''Measured''': ~3
  * '''Measurement method''': REFERENCE TO BE ADDED
Line 130: Line 140:
 * Power at each ISC detection ports when DRFPMI is locked
  * IMC output:
  * REFL:
  * POP p-pol:
  * POP s-pol:
  * POS p-pol:
  * POS s-pol:
  * AS RF:
  * OMC REFL:
  * OMC TRANS:
  * TRX:
  * TRY:
Line 133: Line 155:
  * '''Design''':
  * '''Measured''':
  * '''Measurement method''':
  * '''Design''': 10
  * '''Measured''': 11-12
  * '''Measurement method''': REFERENCE TO BE ADDED
Line 137: Line 159:
 * Power at each ISC detection ports when locked  * Power at each ISC detection ports when PRFPMI is locked
  * IMC output:

Optical Characterization

This page is intended for summarizing the interferometer optical characterization results and discussions. See following pages and documents for iKAGRA and bKAGRA Phase-1 characterization summary:

See following pages for some design parameters

RF Sidebands

Input Mode Cleaner

Arm Cavities

Power Recycling Cavity

Signal Recycling Cavity

  • SRX length
    • Design: 68.2563 m (SRC length = 66.591 m)

    • Measured:

    • Measurement method:

    • Notes:

  • SRY length
    • Design: 64.9265 m

    • Measured:

    • Measurement method:

    • Notes:

  • SRC Gouy phase

Michelson

  • Schnupp asymmetry

Dual Recycled Michelson

  • Power recycling gain
    • Design:

    • Measured: ~3

    • Measurement method: REFERENCE TO BE ADDED

    • Notes:

  • Power at each ISC detection ports when DRFPMI is locked
    • IMC output:
    • REFL:
    • POP p-pol:
    • POP s-pol:
    • POS p-pol:
    • POS s-pol:
    • AS RF:
    • OMC REFL:
    • OMC TRANS:
    • TRX:
    • TRY:

Power Recycled Fabry-Perot Michelson

  • Power recycling gain
    • Design: 10

    • Measured: 11-12

    • Measurement method: REFERENCE TO BE ADDED

    • Notes:

  • Power at each ISC detection ports when PRFPMI is locked
    • IMC output:
    • REFL:
    • POP p-pol:
    • POP s-pol:
    • POS p-pol:
    • POS s-pol:
    • AS RF:
    • OMC REFL:
    • OMC TRANS:
    • TRX:
    • TRY:

Output Mode Cleaner

  • Round-trip length
    • Design:

    • Measured:

    • Measurement method:

    • Notes:

  • Finesse
    • Design:

    • Measured:

    • Measurement method:

    • Notes: