Minute of PEM meeting
- Date: 2019/8/28 17:00-18:00(JST)
- Attend:Yokozawa, TTanaka, Kaihotsu, Mori, kozakai, Washimi, Fujikawa, TTanaka(Niigata), Miyoki, Ohkawa, Kokeyama
- Agenda
- Yokozawa
- Coherent analysis for the FPMI
- Using Bruco, I evaluated the origin of the noise source during the FPmi
- Results are written in klog everyday
- Coherent analysis for the FPMI
- Washimi
- Magnetic field simulation
- Lightning issue
- Discussion about the lightning monitor at 12th Sep.
- monitoring the MAX/MIN magnetometer time series monitoring
- There are lightning candidate around the evening
- Full sampling rate time series monitoring
- correlation between BS oplev was observed only 5th Aug.
- PEM injection T~e-9 but 5thAug is e-6 level
- Need very large coil
- If we use the frame, we can make larger magnetic field
- PSL 44Hz noise hunting
- PSL44Hz
- Can remove this effect. Move the 24V power supply to ouutside of the PSL room. Portable1 acc moved to white laser lack.
- IMMT1 vibration
- ISS cover check
- Check the portable acc, but not detected the serious noise (structure below 40Hz), there seems no large noise structure candidate..
- Kozakai
- Glitch issue
- Omicron was unstable now
- Development the spectrogram tool
- Pick up the requirement to PEM for the JPS meeting
- Glitch issue
- TTanaka
- Preparing for the PEM injection
- PEM injection from analysis building without KAGRA DAQ
- Easy set, easy injection
- Select the PC there is useful injection software, and via USB or something, we can inject remotely if we set the injector to the mine
- Discussion about the future plan this weekend
- Preparing for the PEM injection
- Miyo
- Seismic motion issue if there is topic
- Okutomi
- Shaker injection
- Kaihotsu
- Vacuum pump vibration and comparison between TEAC706 and chromebook
- Using the TEAC706 and KAGRA DA, he checked the vibration of the vacuum pump very carefully
- Actually, there is large 312.5Hz noise. there is 315Hz in the PR2 vacuum pump instead of the 312.5Hz.
- If you read the controller, you can check the rotation rate
- Vacuum pump vibration and comparison between TEAC706 and chromebook
- Fujikawa
- Fujimopy
- Fujikawa+Mori
- Extract the line peak
- Implementing the various noises
- Lockloss with earthquake
- Earthquake vs lockloss
- check by the MCF seismometer
- He found the some earthquake make lockloss
- Investigated the lockloss vs amplitude and source of the earthquake
- Fujimopy
- Yuzurihara
- Yuzu summary page comments
- etc
- Yokozawa