Wide Angle Baffle Cooling Test 2018
| 
 
 | 
 
 | 
 
 | 
 
 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Important note about data
Coil Injection Time
Links to knowledge
- Calculations regarding the back-scattering of different Wide Angle Baffle designs (Zeidler) 
- Personal calc of WAB's stray light noise by Akutsu; due to several uncertain parameters there are several results and no one would know the actual effect! 
- Heat links positions (by Obuchi) fmkopgehpapgdkcc.png 
Task list
Note: some of the tasks can be done in parallel.
- Place/install the WAB suspension. 
- Measure several relative dimensions to get useful information for the future commissioning; concretely speaking: - What....?
 
- Test the saving mode of the WAB suspension. 
- Don't forget to replace the damping magnets to SmCo ones. - What is the requirement to say the damping is sufficient...?
 
- Connect electric cables (with checking) in the IYC chamber. - 3 photosensors; in each photosensor head, two PDs should be connected parallel. Need to calibrate the response of each sensor head... 
- 2 heaters; should be connected parallel?? 
- 1 coil for actuation. 
- 6 thermometers with (extended?) patch cables made of phospher bronze; the extended double twisted pare cables will be delivered on Jan 26 
 
- Connect 4 heat links. 
- Check oplev signals. 
- Connect cables the outside as well. 
- Make a realtime model to monitor. 
- Take interface with Tcam. 
- Make a plan to do leak check before cooling down. 
- Don't forget to add some heat links (Cu?) to the top of the suspension... 
- Remove the Heat-Link block 
Photosensor
The PDs and LED are aligned as shown in the below picture.
 
  
 
The following LEDs and PDs are used; thanks for the information from Fukunaga-kun and Ushiba-kun.
OSEM satellite amp is used to drive.
Pin Assign
- Summary of the pin assign of burndy connector inside the cryostat.
NOTE: The meaning of color for thermometer cable is 
 RED = I- , YELLOW = I+ , GREEN = V- , BLACK = V+  
P22
| Pin Name | Resistance | Equipment | Assign | Comment1 | Comment2 | Connection | 
| A | 75 ohm | Thermometer | V- | Green | Cryocon CH A | ○ | 
| B | 75 ohm | Thermometer | V+ | Black | 
 | ○ | 
| C | 4 ohm | Thermometer | I- | Red | 
 | ○ | 
| D | 4 ohm | Thermometer | I+ | Yellow | 
 | ○ | 
| E | 75 ohm | Thermometer | V- | Green | Cryocon CH B | ○ | 
| F | 75 ohm | Thermometer | V+ | Black | 
 | ○ | 
| G | 4 ohm | Thermometer | I- | Red | 
 | ○ | 
| H | 4 ohm | Thermometer | I+ | Yellow | 
 | ○ | 
| J | 75 ohm | Thermometer | V- | Green | Cryocon CH C | ○ | 
| K | 75 ohm | Thermometer | V+ | Black | 
 | ○ | 
| L | 4 ohm | Thermometer | I- | Red | 
 | ○ | 
| M | 4 ohm | Thermometer | I+ | Yellow | 
 | ○ | 
| N | Blank | 
 | 
 | 
 | 
 | 
 | 
| P | 75 ohm | Thermometer | V- | Green | Cryocon CH D | ○ | 
| R | 75 ohm | Thermometer | V+ | Black | 
 | ○ | 
| S | 4 ohm | Thermometer | I- | Red | 
 | ○ | 
| T | 4 ohm | Thermometer | I+ | Yellow | 
 | ○ | 
| U | 75 ohm | Thermometer | V- | Green | Cryocon CH E | ○ | 
| V | 75 ohm | Thermometer | V+ | Black | 
 | ○ | 
| W | 4 ohm | Thermometer | I- | Red | 
 | ○ | 
| X | 4 ohm | Thermometer | I+ | Yellow | 
 | ○ | 
| Y | Blank | 
 | 
 | 
 | 
 | 
 | 
P23
| Pin Name | Resistance | Equipment | Assign | Comment1 | Comment2 | Connection | 
| A | 75 ohm | Thermometer | V- | Green | Cryocon CH F | ○ | 
| B | 75 ohm | Thermometer | V+ | Black | 
 | ○ | 
| C | 4 ohm | Coil Mag. Act. | I- | 
 | As low resistor line was needed. | ○ | 
| D | 4 ohm | Coil Mag. Act. | I+ | 
 | As low resistor line was needed. | ○ | 
| E | 75 ohm | PS1 PD | V- (K) | 
 | 
 | ○ | 
| F | 75 ohm | PS1 PD | V+ (A) | 
 | 
 | ○ | 
| G | 4 ohm | PS1 LED | I- (K) | 
 | 
 | ○ | 
| H | 4 ohm | PS1 LED | I+ (A) | 
 | 
 | ○ | 
| J | 75 ohm | PS2 PD | V- (K) | 
 | 
 | ○ | 
| K | 75 ohm | PS2 PD | V+ (A) | 
 | 
 | ○ | 
| L | 4 ohm | PS2 LED | I- (K) | 
 | 
 | ○ | 
| M | 4 ohm | PS2 LED | I+ (A) | 
 | 
 | ○ | 
| N | Blank | 
 | 
 | 
 | 
 | 
 | 
| P | 75 ohm | PS3 PD | V- (K) | 
 | 
 | ○ | 
| R | 75 ohm | PS3 PD | V+ (A) | 
 | 
 | ○ | 
| S | 4 ohm | PS3 LED | I- (K) | 
 | 
 | ○ | 
| T | 4 ohm | PS3 LED | I+ (A) | 
 | 
 | ○ | 
| U | 75 ohm | Thermometer | I- | Red | Cryocon CH F | ○ | 
| V | 75 ohm | Thermometer | I+ | Yellow | Cryocon CH F | ○ | 
| W | 4 ohm | Heater | I- | 
 | 
 | ○ | 
| X | 4 ohm | Heater | I+ | 
 | 
 | ○ | 
| Y | Blank | 
 | 
 | 
 | 
 | 
 | 
 
 
DAQ
Real-time computer: k1???
| Slot | Item | Assigned to | S/N | 
| 42 | <blank> | 
 | 
 | 
| 41 | <blank> | 
 | 
 | 
| 40 | <blank> | 
 | 
 | 
| 39 | <blank> | 
 | 
 | 
| 38 | <blank> | 
 | 
 | 
| 37 | <blank> | 
 | 
 | 
| 36 | <blank> | 
 | 
 | 
| 35 | <blank> | 
 | 
 | 
| 34 | <blank> | 
 | 
 | 
| 33 | <blank> | 
 | 
 | 
| 32 | <blank> | 
 | 
 | 
| 31 | <blank> | 
 | 
 | 
| 30 | <blank> | 
 | 
 | 
| 29 | <blank> | 
 | 
 | 
| 28 | <blank> | 
 | 
 | 
| 27 | <blank> | 
 | 
 | 
| 26 | <blank> | 
 | 
 | 
| 25 | <blank> | 
 | 
 | 
| 24 | <blank> | 
 | 
 | 
| 23 | <blank> | 
 | 
 | 
| 22 | <blank> | 
 | 
 | 
| 21 | <blank> | 
 | 
 | 
| 20 | <blank> | 
 | 
 | 
| 19 | <blank> | 
 | 
 | 
| 18 | <blank> | 
 | 
 | 
| 17 | <blank> | 
 | 
 | 
| 16 | <blank> | 
 | 
 | 
| 15 | <blank> | 
 | 
 | 
| 14 | <blank> | 
 | 
 | 
| 13 | <blank> | 
 | 
 | 
| 12 | AI Chassis | DAC0 | SN??? | 
| 11 | AA Chassis | ADC0 | SN??? | 
| 10 | <blank> | 
 | 
 | 
| 9 | I/O Chassis | 
 | SN??? | 
| 8 | <blank> | 
 | 
 | 
| 7 | <blank> | 
 | 
 | 
| 6 | <blank> | 
 | 
 | 
| 5 | <blank> | 
 | 
 | 
| 4 | <blank> | 
 | 
 | 
| 3 | <blank> | 
 | 
 | 
| 2 | <blank> | 
 | 
 | 
| 1 | <blank> | 
 | 
 | 
| 0 | <blank> | 
 | 
 | 
OpLev
- Optical lever installed at the +X/-Y side of the IYC
- Path-length inside the chamber (viewport to viewport): ~3000 mm
 
WAB Alignment
- Alignment done with 2 laser-leveler inside the IYC
- Regarding the used terms: - As the baffle is placed along the Y-arm, +Y and -Y refer to both long ends of the baffle toward the Y-end and toward the center area, respectively!
- The same holds true for the horizontal dimension in x-arm direction (+X means toward X-end and -X toward center/PSL)
 
Yaw
- laser-leveler needs to be close to the -Y main-beam entrance in a height 300 - 400 mm above the optical table
- Vertical laser line to be adjusted to the main-beam indication on the optical table and the inner wall of the cryostat
- Checked the line by using the slids of the flanges on top the WAB (each slid is ~3.5 mm broad)
- For both slids, the laser line is mismatched (not centered) toward -X by ~0.5 mm (@ +Y slid) and ~1 mm (@ -Y slid)
That means that there is probably also a -X offset!
- The flanges have a distance of 170 mm which indicates a maximum yaw-tilting of ~0.1 degrees
Note: taking the distances is very hard as the position of the slids is not easy to access
Pitch
- laser-leveler outside the cryostat (but inside IYC) set to give a horizontal line along length side of the baffle
- Found a slight mismatch for the outer screw holding the OpLev mirrors in +Y,+X and -Y,+X 
- Difference is ~1 mm (+Z @ -Y, so -Y-side is higher than +Y-side)
- With approx. 500 mm distance between both screws, the maximum pitch is ~0.1 degrees
Roll
- Took the height of the OpLev mirror screw-holes (@ -Y,+X and -Y,-X) with a vertical ruler and the horizontal line of the laser-leveler outside the cryostat 
- Measured a difference of 2.5 mm which makes it a roll of 0.55 degrees (radius of baffle ~130 mm); -Y,-X is higher!
- Then took the height of baffle-wings which are mounted on the flanges and could not measure any difference (within an error of 0.5 mm)
- Hence, only the baffle is slightly rolled within the flanges!