Wide Angle Baffle Cooling Test 2018
|
|
|
|
Important note about data
Coil Injection Time
Links to knowledge
Calculations regarding the back-scattering of different Wide Angle Baffle designs (Zeidler)
Personal calc of WAB's stray light noise by Akutsu; due to several uncertain parameters there are several results and no one would know the actual effect!
Heat links positions (by Obuchi) fmkopgehpapgdkcc.png
Task list
Note: some of the tasks can be done in parallel.
Place/install the WAB suspension.
Measure several relative dimensions to get useful information for the future commissioning; concretely speaking:
- What....?
Test the saving mode of the WAB suspension.
Don't forget to replace the damping magnets to SmCo ones.
- What is the requirement to say the damping is sufficient...?
Connect electric cables (with checking) in the IYC chamber.
3 photosensors; in each photosensor head, two PDs should be connected parallel. Need to calibrate the response of each sensor head...
2 heaters; should be connected parallel??
1 coil for actuation.
6 thermometers with (extended?) patch cables made of phospher bronze; the extended double twisted pare cables will be delivered on Jan 26
Connect 4 heat links.
Check oplev signals.
Connect cables the outside as well.
Make a realtime model to monitor.
Take interface with Tcam.
Make a plan to do leak check before cooling down.
Don't forget to add some heat links (Cu?) to the top of the suspension...
Remove the Heat-Link block
Photosensor
The PDs and LED are aligned as shown in the below picture.
The following LEDs and PDs are used; thanks for the information from Fukunaga-kun and Ushiba-kun.
OSEM satellite amp is used to drive.
Pin Assign
- Summary of the pin assign of burndy connector inside the cryostat.
NOTE: The meaning of color for thermometer cable is
RED = I- , YELLOW = I+ , GREEN = V- , BLACK = V+
P22
Pin Name |
Resistance |
Equipment |
Assign |
Comment1 |
Comment2 |
Connection |
A |
75 ohm |
Thermometer |
V- |
Green |
Cryocon CH A |
○ |
B |
75 ohm |
Thermometer |
V+ |
Black |
|
○ |
C |
4 ohm |
Thermometer |
I- |
Red |
|
○ |
D |
4 ohm |
Thermometer |
I+ |
Yellow |
|
○ |
E |
75 ohm |
Thermometer |
V- |
Green |
Cryocon CH B |
○ |
F |
75 ohm |
Thermometer |
V+ |
Black |
|
○ |
G |
4 ohm |
Thermometer |
I- |
Red |
|
○ |
H |
4 ohm |
Thermometer |
I+ |
Yellow |
|
○ |
J |
75 ohm |
Thermometer |
V- |
Green |
Cryocon CH C |
○ |
K |
75 ohm |
Thermometer |
V+ |
Black |
|
○ |
L |
4 ohm |
Thermometer |
I- |
Red |
|
○ |
M |
4 ohm |
Thermometer |
I+ |
Yellow |
|
○ |
N |
Blank |
|
|
|
|
|
P |
75 ohm |
Thermometer |
V- |
Green |
Cryocon CH D |
○ |
R |
75 ohm |
Thermometer |
V+ |
Black |
|
○ |
S |
4 ohm |
Thermometer |
I- |
Red |
|
○ |
T |
4 ohm |
Thermometer |
I+ |
Yellow |
|
○ |
U |
75 ohm |
Thermometer |
V- |
Green |
Cryocon CH E |
○ |
V |
75 ohm |
Thermometer |
V+ |
Black |
|
○ |
W |
4 ohm |
Thermometer |
I- |
Red |
|
○ |
X |
4 ohm |
Thermometer |
I+ |
Yellow |
|
○ |
Y |
Blank |
|
|
|
|
|
P23
Pin Name |
Resistance |
Equipment |
Assign |
Comment1 |
Comment2 |
Connection |
A |
75 ohm |
Thermometer |
V- |
Green |
Cryocon CH F |
○ |
B |
75 ohm |
Thermometer |
V+ |
Black |
|
○ |
C |
4 ohm |
Coil Mag. Act. |
I- |
|
As low resistor line was needed. |
○ |
D |
4 ohm |
Coil Mag. Act. |
I+ |
|
As low resistor line was needed. |
○ |
E |
75 ohm |
PS1 PD |
V- (K) |
|
|
○ |
F |
75 ohm |
PS1 PD |
V+ (A) |
|
|
○ |
G |
4 ohm |
PS1 LED |
I- (K) |
|
|
○ |
H |
4 ohm |
PS1 LED |
I+ (A) |
|
|
○ |
J |
75 ohm |
PS2 PD |
V- (K) |
|
|
○ |
K |
75 ohm |
PS2 PD |
V+ (A) |
|
|
○ |
L |
4 ohm |
PS2 LED |
I- (K) |
|
|
○ |
M |
4 ohm |
PS2 LED |
I+ (A) |
|
|
○ |
N |
Blank |
|
|
|
|
|
P |
75 ohm |
PS3 PD |
V- (K) |
|
|
○ |
R |
75 ohm |
PS3 PD |
V+ (A) |
|
|
○ |
S |
4 ohm |
PS3 LED |
I- (K) |
|
|
○ |
T |
4 ohm |
PS3 LED |
I+ (A) |
|
|
○ |
U |
75 ohm |
Thermometer |
I- |
Red |
Cryocon CH F |
○ |
V |
75 ohm |
Thermometer |
I+ |
Yellow |
Cryocon CH F |
○ |
W |
4 ohm |
Heater |
I- |
|
|
○ |
X |
4 ohm |
Heater |
I+ |
|
|
○ |
Y |
Blank |
|
|
|
|
|
DAQ
Real-time computer: k1???
Slot |
Item |
Assigned to |
S/N |
42 |
<blank> |
|
|
41 |
<blank> |
|
|
40 |
<blank> |
|
|
39 |
<blank> |
|
|
38 |
<blank> |
|
|
37 |
<blank> |
|
|
36 |
<blank> |
|
|
35 |
<blank> |
|
|
34 |
<blank> |
|
|
33 |
<blank> |
|
|
32 |
<blank> |
|
|
31 |
<blank> |
|
|
30 |
<blank> |
|
|
29 |
<blank> |
|
|
28 |
<blank> |
|
|
27 |
<blank> |
|
|
26 |
<blank> |
|
|
25 |
<blank> |
|
|
24 |
<blank> |
|
|
23 |
<blank> |
|
|
22 |
<blank> |
|
|
21 |
<blank> |
|
|
20 |
<blank> |
|
|
19 |
<blank> |
|
|
18 |
<blank> |
|
|
17 |
<blank> |
|
|
16 |
<blank> |
|
|
15 |
<blank> |
|
|
14 |
<blank> |
|
|
13 |
<blank> |
|
|
12 |
AI Chassis |
DAC0 |
SN??? |
11 |
AA Chassis |
ADC0 |
SN??? |
10 |
<blank> |
|
|
9 |
I/O Chassis |
|
SN??? |
8 |
<blank> |
|
|
7 |
<blank> |
|
|
6 |
<blank> |
|
|
5 |
<blank> |
|
|
4 |
<blank> |
|
|
3 |
<blank> |
|
|
2 |
<blank> |
|
|
1 |
<blank> |
|
|
0 |
<blank> |
|
|
OpLev
- Optical lever installed at the +X/-Y side of the IYC
- Path-length inside the chamber (viewport to viewport): ~3000 mm
WAB Alignment
- Alignment done with 2 laser-leveler inside the IYC
- Regarding the used terms:
- As the baffle is placed along the Y-arm, +Y and -Y refer to both long ends of the baffle toward the Y-end and toward the center area, respectively!
- The same holds true for the horizontal dimension in x-arm direction (+X means toward X-end and -X toward center/PSL)
Yaw
- laser-leveler needs to be close to the -Y main-beam entrance in a height 300 - 400 mm above the optical table
- Vertical laser line to be adjusted to the main-beam indication on the optical table and the inner wall of the cryostat
- Checked the line by using the slids of the flanges on top the WAB (each slid is ~3.5 mm broad)
- For both slids, the laser line is mismatched (not centered) toward -X by ~0.5 mm (@ +Y slid) and ~1 mm (@ -Y slid)
That means that there is probably also a -X offset!
- The flanges have a distance of 170 mm which indicates a maximum yaw-tilting of ~0.1 degrees
Note: taking the distances is very hard as the position of the slids is not easy to access
Pitch
- laser-leveler outside the cryostat (but inside IYC) set to give a horizontal line along length side of the baffle
Found a slight mismatch for the outer screw holding the OpLev mirrors in +Y,+X and -Y,+X
- Difference is ~1 mm (+Z @ -Y, so -Y-side is higher than +Y-side)
- With approx. 500 mm distance between both screws, the maximum pitch is ~0.1 degrees
Roll
Took the height of the OpLev mirror screw-holes (@ -Y,+X and -Y,-X) with a vertical ruler and the horizontal line of the laser-leveler outside the cryostat
- Measured a difference of 2.5 mm which makes it a roll of 0.55 degrees (radius of baffle ~130 mm); -Y,-X is higher!
- Then took the height of baffle-wings which are mounted on the flanges and could not measure any difference (within an error of 0.5 mm)
- Hence, only the baffle is slightly rolled within the flanges!